Dräger Atlan A100 Instructions For Use Manual page 149

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Using the O
– In manual ventilation, pressing the O
– In mechanical ventilation, permanently pressing the O
8.3.3
Using the vaporizer
● Operate the vaporizer according to its instructions for use.
8.3.4
Flow correction
The gas composition can affect the measurement accuracy of the flow
measurement.
On devices equipped with an integrated patient-gas measurement module, the flow
measurement is corrected automatically.
On devices equipped with inspiratory O
measurement can be ensured by manual or automatic flow correction.
Prerequisites for automatic flow correction on devices equipped with inspiratory O
measurement:
– The anesthetic gas monitor (Dräger Vamos or Dräger Scio) is connected to the
COM 2 serial port.
– The COM 2 serial port is activated in the system setup, see the table on
page 220.
1. Touch area (1); a further dialog (2) will open.
CAUTION
Risk of inaccurate measured flow values
If the flow is not set correctly, the resulting incorrect measured flow values can lead
to incorrect measured values for the tidal volume.
► Make the settings for the flow correction carefully.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
flush has the following effects:
2
to the APL level.
slight rise of the PEEP level. However, this rise has no effect on the peak
pressure.
4
1
flush results in a rapid rise in pressure
2
measurement, the accuracy of the flow
2
2
3
5
Operation
flush may result in a
2
2
149