The following table gives an overview of the integrated safety functions which come
into effect if problems occur during operation:
Fault
Leakage
Mains power supply
failure
Mains power supply
failure and battery
discharged
Failure of the central
gas supply system
Complete failure of
the gas supply
Flow measurement
failure
Failure of fresh-gas
delivery (electroni-
cally controlled gas
mixer only)
Ventilator failure
54
Safety function
Pressure regulation and PEEP compensation
– In mechanical ventilation modes, the set PEEP is
actively adjusted even in the case of small leaks.
– In pressure-controlled modes, the pressure is con-
trolled during the entire respiratory cycle.
Uninterruptible power supply provided by internal battery
– Device runtime in battery operation (see "Technical
data", page 272)
– Automatic deactivation of the breathing system warmer
to increase battery runtime
– Manual ventilation and spontaneous breathing are
available
– Emergency O
mixer)
– Fresh-gas delivery (mechanically controlled gas mix-
ers)
– Delivery of anesthetic agents via connected vaporizers
– Use of the connected gas cylinders
– Mechanical ventilation with ambient air possible (the
hose with the breathing bag will have to be removed for
this)
– No anesthetic agent delivery from the connected
vaporizers possible; switch to intravenous anesthetic
agent required
– Mechanical ventilation can be continued.
– Restrictions with regard to displayed measured values,
measurement accuracies, and when triggering manda-
tory breaths are possible.
– Emergency O
(electronically controlled gas mixer)", page 50)
– Anesthetic agent delivery from connected vaporizers
possible
– All ventilation modes are available.
Alternatively:
– Backup manual mode
– Manual ventilation or spontaneous breathing possible
– Fresh-gas delivery available
– Anesthetic agent delivery from connected vaporizers
possible
Alternatively:
– Backup manual mode
|