Follow the instructions for use for the particular soda lime.
4.9.3
Water trap
● Empty or replace the water trap according to its instructions for use.
4.9.4
Connecting the sample line
WARNING
Risk of underpressure in the lungs
In the following cases, the sample gas flow can immediately cause negative
pressure in the lungs, particularly in neonates or pediatric patients:
– Filters or hoses are clogged.
– The sample gas is being extracted between the patient and an occluded
component.
Ensure the following when ventilating pediatric patients and neonates:
► When using a fine-pored filter close to the patient:
– Do not connect the sample line between the tube and the filter. Instead,
► When using an HME and a sample line port on a tube adapter close to the
patient:
– Use only HMEs that are listed in the list of accessories.
– Set the alarm limits for MV low and Paw to suitable values.
– Observe the following information: "CO
WARNING
Risk due to incorrectly connected sample line
If the sample line is connected to the wrong connectors, e.g., connectors on
infusion pumps, fluids may be drawn in instead of sample gas. As a result, the gas
measurement cannot not display correct values or the device can become
damaged.
► When connecting the sample line, take care that it is correctly connected.
WARNING
Risk of incorrect measured gas values
Blocked water traps or blocked sample lines prevent correct gas measurement. As
a result, incorrect measured gas values could be displayed.
► Only use Dräger sample lines.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
connect it directly to the filter or the Y-piece.
and neonates", page 321.
Assembly and preparation
measurement with pediatric patients
2