Operation
Mechanically controlled gas mixer with electronic flow measurement:
● Open the flow control valves. The device switches automatically to the Manual /
Spontaneous mode. Take care that the quick start does not happen
unintentionally.
Electronically controlled gas mixer:
1. Observe the fresh-gas settings (3).
2. Touch the Quick start Man/Spon button (1).
Normal start with customized settings
1. Optionally perform one of the following steps:
● Touch the Start... button (2).
● Touch the screen in the monitoring area (4).
● Squeeze the breathing bag twice.
2. Adjust the patient data and ventilation settings.
Starting when time is limited
When time is limited, adjusting the patient data can be omitted. Start the therapy as
follows:
1. Touch the screen.
2. Check the displayed start values.
3. Press the rotary knob. The therapy starts.
4. Adjust the patient data as soon as possible.
8.2.1
Loading the patient data
There are two possibilities for loading the patient data:
– Starting a new case (1)
– Continuing a case (2)
Depending on the selected patient category, different patient data (e.g., weight,
age) are displayed in area (3).
144
1
Start
Select patient and check data
3
Select mode
Instructions for use
2
|
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n