Dräger Atlan A100 Instructions For Use Manual page 165

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

8.6.2
Econometer
Prerequisite: The device has the "advanced gas monitoring" software option.
During operation, the device monitors the breathing bag for sufficient filling.
Surplus
Efficient
Deficit
Econometer
The bar graph indicates whether the device is supplied with sufficient fresh gas.
Range
Surplus
Efficient
Deficit
An insufficiently filled breathing bag can trigger the Fresh gas low or leakage or
Ambient air inlet activated alarms, for example.
8.6.3
Stopwatch
Using the stopwatch
1. Touch the Start button (1) to start.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Color
Yellow
Green
Red
Stopwatch
Meaning
Indication of an opportunity to save fresh
gas and, therefore, volatile anesthetic
agents
– No action necessary
– Breathing bag sufficiently filled
– Sufficient reserve capacity available
– Insufficient fresh-gas supply
– Check the filling of the breathing bag. If
necessary, fill up the breathing bag,
e.g., with the O
flush.
2
1
Operation
165