8.4.3
Monitoring operation mode
Prerequisites:
– The device is equipped with the "integrated patient-gas measurement module".
– The Pause operation mode is switched off in the system setup (see "Vertical tab
"General"", page 214).
In Monitoring operation mode, breathing gas measurement can be used to monitor
the patient's spontaneous breathing activity without distracting ventilation alarms.
This mode is useful during regional anesthesia or after extubation. To do this,
connect the sample line of the PGM to the breathing mask used. At the same time,
O
insufflation can be applied.
2
In this mode, there is no automatic gas delivery. The breathing system is
depressurized so that manual ventilation is not possible. The gas concentration
measurement is active and waiting for respiratory phases. With the detection of
respiratory phases, apnea monitoring is automatically activated.
8.4.3.1
Activating monitoring
Prerequisite: The device is in standby or Manual / Spontaneous mode.
● Start the Monitoring operation mode (1).
8.4.3.2
Ending monitoring
● After standby or switching to a ventilation mode.
|