WARNING
Risk of fire
In combination with oxygen or nitrous oxide, ignition sources such as
electrosurgical devices and laser surgical devices can cause fires.
► If ignition sources are present, do not open the flow control valve on the O
flowmeter. Leave the flow control valve completely closed.
2. Open the flow control valve (1) and set the desired O
the O
3. Monitor the flow on the total flow tube (2).
4. Set the anesthetic gas concentration on the vaporizer.
5. Ventilate the patient manually.
6. Turn on the device.
7. As soon as the Standby page is displayed, start the therapy, see page 143.
6.7.3
Electronically controlled gas mixer
1. Adjust the APL valve.
2. Set the O
WARNING
Risk of fire
In combination with oxygen or nitrous oxide, ignition sources such as
electrosurgical devices and laser surgical devices can cause fires.
► If ignition sources are present, do not open the flow control valve on the O
flowmeter. Leave the flow control valve completely closed.
3. Open the flow control valve (2) and set the desired O
the O
4. Monitor the set flow on the O
5. Set the anesthetic gas concentration on the vaporizer.
6. Ventilate the patient manually.
7. Turn on the device.
8. As soon as the Standby page is displayed, start the therapy, see page 143.
9. Set the O
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
+ key (3) to quickly fill the breathing bag.
2
4
3
switch (3) to the Add. O
2
+ key (4) to quickly fill the breathing bag.
2
switch (3) to Aux. O
2
1
2
position.
2
flowmeter (1).
2
.
2
Getting started
2
flow. If necessary, press
2
2
flow. If necessary, press
2
109

Advertisement

Table of Contents
loading

Table of Contents