4.3.3
Establishing the mains power supply
The mains voltage must correspond to the voltage range indicated on the rating
plate on the rear of the device.
To protect from inadvertent disconnection of the power cable, the power inlet of the
device is secured with a guard plate.
WARNING
Risk due to incorrect mains voltage or missing protective ground
If the device is connected to a power socket with incorrect mains voltage or a power
socket without a protective ground, an electric shock may occur.
► Connect the device only to power sockets with correct mains voltage and a
protective ground.
NOTICE
Risk of electrical overload
If the device is connected to an additional power socket strip, this may cause an
electrical overload.
► Do not connect the device to power socket strips.
► Do not connect additional power socket strips to the auxiliary power sockets on
the device, but rather to separate power sockets on the wall.
1. Plug the power plug into the power socket.
device can be interrupted quickly in the event of a device malfunction.
2. Using the information displayed on the gas mixer unit, check that power is
supplied.
3. Turn on the device.
4.3.4
Charging the battery
The internal battery will automatically start charging as soon as the device is
connected to the mains power supply.
4.3.5
Connecting other devices to the auxiliary power sockets
An illustration of the auxiliary power sockets can be found on page 33.
● Connect the power cable of the other device to an auxiliary power socket.
Make sure that the maximum current consumption of the other devices does not
exceed the permissible value.
WARNING
Risk of fire
Components that can heat up or cause sparks, such as electrical items, batteries,
or power supply units, are unable to cool down in enclosed storage locations and
can cause a fire.
► Do not keep electrical items in the drawers or in the storage compartment at the
rear of the device.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
The power plug must be readily accessible so that the power supply to the
Assembly and preparation
73