Dräger Atlan A100 Instructions For Use Manual page 158

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Operation
Relevant waveforms, trends, or measured values are displayed in area (1) of the
dialog window. The parameters for the particular maneuver are displayed and set in
area (2).
8.5.2
Inspiration Hold, Expiration Hold
The device provides functions for a breath to be initiated or extended, or for
expiration to be extended.
This can be useful in situations where the lungs of the patient are not supposed to
move, e.g., during use of imaging techniques.
8.5.2.1
Manual Inspiration / Inspiration Hold
This maneuver is available in the volume-controlled modes, the pressure-controlled
modes, and in the CPAP / PSV mode, and offers the following options:
– In the expiratory phase between 2 mandatory breaths, a breath can be manually
triggered and held. The ventilation pattern of the manually triggered breath
corresponds to the ventilation pattern for the active ventilation mode.
– A mandatory breath can be extended.
158
1
Procedures
2
Procedures
Instructions for use
|
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n