Safety-related information
Instructions for use for accessories
If accessories or connected devices are used contrary to the information in the
associated instructions for use, this may lead to user errors, incorrect use, or
incorrect reprocessing. Personal injury and property damage may occur as a
consequence.
► Follow the instructions for use for all accessories, e.g.:
– Water traps
– Flow sensors
– CLIC adapter
– CLIC absorber
– Soda lime
– Breathing hoses
– Masks
– Filter
– Bronchial suction
– Vaporizer
– Manual resuscitator
– AGSS terminal unit
2.9.5
Color codes and labels
Arrangement and display of gases
In some countries, the arrangement and display of the gases on the gas mixer and
on the screen may deviate from the illustrations shown in this document.
► Always pay attention to the respective color codes and labels.
2.9.6
Device
Penetrating liquid
Penetrating liquid may cause the following:
– Damage to the device
– Electric shock
– Device malfunctions
Personal injury and property damage may occur as a consequence.
► Ensure that no liquid, e.g., disinfectant, penetrates the device.
► Do not place any containers containing liquids above or on the device.
► Do not transport infusion bags or other containers with liquids with the device.
► If the device was exposed to moisture or liquids during transport, wipe it/them off
after transport. Before connecting the device to the power supply, ensure
sufficient drying time.
► Only use the device on level surfaces without inclination.
14
|