Backup manual mode with an electronically controlled gas mixer
1. Open the flap (1).
2. Activate the backup manual switch (2). Follow the instructions on the product
label (3) or the screen (4).
3. Set the O
4. Open the flow control valve on the emergency O
O
5. Ventilate the patient manually.
a. During mechanical ventilation or in Man/Spon ventilation mode:
b. When using the external fresh-gas outlet:
6. Check the vaporizer setting.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
4
switch to Add. O
2
flow. The set O
flow constitutes the total fresh-gas flow.
2
2
Ventilate manually with the breathing bag.
Ventilate manually with the breathing bag on the non-rebreathing system.
.
2
delivery and set an adequate
2