The automatic test consists of a pretest (the initialization), which may require
manual actions by the user, and a fully automatic main test.
On devices with an O
required. If more than 7 days have passed since the last calibration of the O
sensor, an O
On devices with an integrated patient-gas measurement module, automatic
calibrations are performed regularly during operation.
7.4.1.1
Execution
WARNING
Risk of patient injury
During the system test, the device is pressurized.
► To prevent patient injury, do not perform the system test and leakage test if a
patient is connected.
1. Touch the button for the system test.
2. Complete the tabular checklist (1).
It is possible to switch from the tabular checklist to the walk-through mode, if
required. To do this, touch the button (2). For further information see: "Checklist
in the walk-through mode", page 125.
3. If the O
Calibrate button (3).
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
sensor, the system test checks whether an O
2
calibration will be automatically performed.
2
System test - checklist
2
sensor is to be calibrated despite a valid calibration, touch the
2
1
4
5
Tests
calibration is
2
2
3