Troubleshooting
Priority
Alarm
!
100
Water trap expired
!
170
Water trap full
!!! 255
Wrong patient connection? The patient is connected to
!! 80
xMAC low
!
170
xMAC low
266
Cause
Accessory has been used
too long.
The water trap of the gas
measurement is full.
Sample line is occluded.
the internal breathing sys-
tem, but there is no oxygen
delivery.
Breathing activity has been
detected at an inactive
breathing system.
Inspiratory and expiratory
gas concentrations are lower
than the automatically calcu-
lated limit.
Inspiratory and expiratory
gas concentrations are lower
than the automatically calcu-
lated limit.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Remedy
Replace the accessory if nec-
essary.
Use "ALARM RESET" to
reset alarm.
Check water trap.
Check sample line, water
trap, and patient-side filter.
Change to Man/Spon mode
and ventilate the patient
manually.
Check if the patient is con-
nected to the correct breath-
ing system.
Check patient condition.
Check filling level. Refill if
necessary.
Check vaporizer setting.
Check the breathing system
and the breathing bag for
leakages.
If the current xMAC is accept-
able, use "ALARM RESET"
to reset the alarm.
Check patient condition.
Check filling level. Refill if
necessary.
Check vaporizer setting.
Check the breathing system
and the breathing bag for
leakages.
If the current xMAC is accept-
able, use "ALARM RESET"
to reset the alarm.