Configuring Input Channels
Measuring voltage
This section describes how to configure settings on the individual settings screen when measuring
voltage.
You can use
Applicable modules: U8550, U8551, U8552, U8553, U8554, LR8530, LR8531, LR8532, LR8533,
>
1
Select the module (Unit) and channel to configure and select the check box.
Measurement will not be performed for channels whose check boxes are not selected.
2
Select the waveform display color.
×
(OFF), 24 colors
Select
[×]
screen.
3
Set the input type to [Voltage].
For the U8553 and LR8533, the setting cannot be changed from [Voltage].
4
Under [Range], select the measurement range as appropriate for the measurement target.
The measurable range of the selected range will be displayed.
(For the U8550, U8551, U8552, LR8530, LR8531, or LR8532)
10 mV
(For the U8553 or LR8533 High Speed Voltage Unit)
100 mV
(For the U8554 or LR8534 Strain Unit)
1 mV
,