LR5001 Instruction Manual HUMIDITY LOGGER Read carefully before use. Keep for future reference. When using the instrument for the Troubleshooting first time Maintenance and Service p.91 Part Names/Functions p.12 and Display Indicators Troubleshooting p.92 Settings List p.27...
Installing the Logger ..........40 Starting and Stopping Recording ......42 Confirming Currently Measured Values and Data Recording ............45 Automatically Importing (Saving) Recorded Data to a Computer, and Graph Display ......45 HIOKI LR5001B980-05 LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at:...
Introduction Introduction Thank you for purchasing the HIOKI "Model LR5001 Humidity Logger." To obtain maximum performance from the instrument, please read this manual first, and keep it handy for future reference. The latest edition of the instruction manual The contents of this manual are subject to change, for example as a result of product improvements or changes to specifications.
When you receive the instrument, inspect it carefully to ensure that no damage occurred during shipping. In particular, check the accessories, panel switches, and connectors. If damage is evident, or if it fails to operate according to the specifica- tions, contact your dealer or Hioki representative. Quantities in parentheses ( ). LR5001...
Page 7
The options listed below are available for the instrument. To order an option, please contact your authorized Hioki distributor or reseller. Options are subject to change. Please check Hioki's website for the latest information. LR5091 Communication Adapter(1) LR9504 Humidity Sensor (Includes LR5000 Utility Program* CD [PC (Approx.
Indicates that incorrect operation presents a possibility of injury to the user or damage to the instrument. Indicates advisory items related to performance or correct operation of the instrument HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at:...
Exposed high fields humidity or condensa- Near electromagnetic tion radiators Near induction heat- ing systems (e.g., high-frequency Subject to vibration induction heating sys- tems and IH cooking utensils) HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 10
LR5091 Communication Adapter and LR5092-20 Data Collector. Avoiding Instrument Damage To avoid damage to the instrument, protect it from physical shock when transporting and handling. Be especially careful to avoid physical shock from dropping. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at:...
Page 11
Never use abrasives or sol- vent cleaners. • Hioki shall not be held liable for any problems with a computer system that arises from the use of this CD, or for any problem related to the purchase of a Hioki product.
Example Case: Record warehouse temperature and humidity at 10-minute intervals for one month, and store the data on a computer. Required Items: Quantities in parentheses ( ). LR5001 (1) LR6 alkaline battery (1) LR9504 Humidity Sensor LR5091 Communication Adapter(1) Computer (1)
Page 13
(This setting provides one-time measurement: recording stops when memory becomes full.) "Stop Method Setting (for when See: memory becomes full)" (p.30) The setting can be made also from the LR5000 Utility Program.) (p.35) HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 14
"4.5" (p.45) ÅiThe data is automatically saved when imported to the computer. By default, it is also automatically displayed in a graph.) Print recorded data as needed. "4.8" (p.60) See: HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Changes setting values on the Settings display. LR5091 Communication Adapter IR Port (p.45) Communicates with the logger. USB Port (p.32) Connect a USB cable here to communicate with a computer. (Mini-B receptacle) HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 17
* Setting is available from the LR5000 Utility Program or via the LR5092-20 Data Col- lector. "3.3 Making Settings from the LR5000 Utility Program" (p.32), LR5092-20 Data See: Collector Instruction Manual HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at:...
Channel 2 second. maximum value • The maximum and minimum values are Channel 2 not displayed when the recorded data minimum value count is 0. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 20
(although they can still be displayed). • Settings cannot be changed while recording. However, settings can still be displayed by pressing the button from the Measurement display. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
1. All segments 4. Year Setting display 2. Model name 3. Firmware version • When the battery indicator appears, settings cannot be changed (although they can still be displayed). HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 22
When the logger will not be used for long time To avoid corrosion and damage to this instrument from battery leakage, remove the batteries from the instrument if it is to be stored for a long time (1 week). HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 23
The cover will not close correctly if there are any gaps. Never attempt to force the battery cover closed when not aligned properly. Doing so could cause damage. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
• To avoid damage to the instrument, do not apply voltage to sensor jacks. Connection Method Required Items: Hioki LR9501 to LR9504 Temp/Humidity Sensor Align the triangle on the plug with the one in front of the sensor jacks, and insert the plug securely.
Set the included CD to the CD-ROM drive. Click [Start] to display the application list. Click [Windows System] [File Explorer] to start Explorer. Click [This PC], and then, double-click [CD Drive (D)] drive. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 26
CD-R. In such a case, download the executable program from the “Drivers, Firmware, Soft- ware” page of Hioki’s website, and then install it again. • The data logger series LR5000 programs consists of LR5000 Utility Pro- gram and LR5091/LR5092 Device Driver, both of which need to be installed.
Page 27
Download the latest version of the LR5000 Utility Program from our website (http:// www.hioki.com). Follow the procedure on the download page to install the latest ver- sion. (The old version is uninstalled automatically.) HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at:...
Page 28
Help. Logger data import Displays Setting Screens: Display Data Display Data Viewing screens Sorting screens Logger settings Setting Screens (p.33) Make and export logger settings. Example: Logger settings HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 29
Example: Logger import screen Data Viewing Screens (p.58) View imported data on these screens. Select a file to view, as a graph or table. Example: Screens for viewing the latest data HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at:...
Page 30
You can copy, delete, move, combine, and extract data. Example: Data Copy screen Option Screens (p.77) Make advanced settings on these screens. You can specify the data importing method. Example: Import Method Setting screen HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
(p.34) specified logger. Enter a comment for the Channel Comment specified measurement (p.34) channel. Select the recording start Recording Start method. (The start time can (p.35) Method be specified.) HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 32
Use to scale measured val- Scaling ues to display as adjusted (p.36) values. Set upper and lower thresh- old values to display the Alarm Thresholds (p.37) alarm indicator [AL] on the logger. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
(The month setting starts blinking.) Repeat this procedure to set the month, day, hour, and minute. Press the button to accept the setting. (The stop method setting is displayed.) HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 34
When memory becomes full during endless recording, the recorded data count (equal to the memory capacity) remains constant. (instantaneous value recording display) (statistical value recording display) HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at:...
Page 35
Power save is enabled. Power save is disabled (the display remains visible). The Auto Power Save feature consumes a small amount of current "Appendix 3 Battery Life Approximation" (p.A3) See: HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at:...
If the main screen is not displayed on the computer, click the icon in the task tray (notification area), and click [Show Main Click Screen]. The main screen appears. Click HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 37
[Import Settings] button at the upper right of screen. (A dialog appears.) 2. Click the [Import Settings to Computer] button. (The logger's settings are now reflected in the program.) HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 38
The following characters are not allowed: \, /, :, *, ?, ", <, >, and |. Settings on the [Recording Method] The Auto Power Save feature consumes a small amount of current HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at:...
Page 39
Statistical recording results in shorter battery life. "Appendix 3 Battery Life Approxima- See: tion" (p.A3) Statistical recording cannot be selected when the recording interval is set to one second. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 40
Select this check box to enable scaling. Specify by example, or Specify by A/B Clicking this changes the setting options. Make set- tings on either tab. (The settings are ap- plied to the other tab.) HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 41
Note: Alarm judgment is performed at every recording interval during instantaneous recording, and once per second during statistical recording. Note: Alarm judgment is performed using measurement values with a larger number of digits than the values (4 digits) indicated in the LR5001 display. Note: The [AL] indicator appears when the measured value is out of range (OF/UF dis- played), and when a sensor anomaly occurs (- - - - displayed).
Page 42
(1) The logger measures 0.2°C and the reference device measures 0.0°C. Since only one point is known, set the slope to "1" and enter the offset only. [°C] Slope (coefficient A) [°C] -0.2 Offset (coefficient B) HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
TV, or electronic wristwatch. Using the Stand Required Items: Stand(Accessory) Strap/stand attach- Attach the stand to the strap/stand attachment ment hole hole. Stand up the logger. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 45
Required Items: Z5004 (Option) Attach the Z5004 to the strap/stand attach- ment hole. (feed the strap through a, b, and c) Attach the magnet to the wall (ferrous mate- rial). Magnet HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Recorded Data to a Computer, and Graph Display" (p.45) [REC] disappears when recording stops. If the stop method is set to [OFF] (one-time recording), recording stops automatically when memory becomes full.(p.30) HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 47
If the recording interval is 10 minute and recording is started at 12:00:00, the first data will be from 12:10:00. Power Control (Approx. 30 12:00:00 12:10:00 Seconds) Recording Start Operation Actual Recording Start HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 48
Recording starts. In the case of instantaneous value recording, the number of data soon becomes 1. In the case of statistical value recording, the number of data becomes 1 at 12:20:00. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at:...
Data recorded in the logger can be imported to the computer. Install the LR5000 Utility Program on the computer beforehand. (p.21) Required Items: LR5001, LR5091 Communication Adapter (or LR5092-20 Data Col- lector), USB cable, Computer Plug the USB cable into the USB port on the LR5091 (or LR5092-20), and into a USB port on the computer.
Page 50
How is recorded data saved? Recorded data is automatically saved when imported to a computer. The save destination and file name are specified as a basic setting on the Options screen. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at:...
Page 51
Displays the [Graph Set- tings] dialog box when a graph is displayed. (p.50) Click buttons switch between graph and table displays. The graph or table is displayed. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 52
Converts data outside of the upper and lower thresh- OVER Data Revision old settings to specified values, and saves as new data. (p.70) Help Displays the help file. Help Displays LR5000 Utility Program version informa- Version tion. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 53
[Item settings] tab in the [Statistical Information and Item Settings] dialog box (p.52) How can graph details be set? Detailed settings are available in the [Graph Settings] dialog box. (p.50) HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 54
Click [Execute] to apply the settings. Specifies cursor positions. Click [Exe- cute] to apply the settings. Specifies the graph start position (time). Click [Execute] to apply the settings. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 55
Upper and lower thresholds can be dis- played as solid lines on the graph, or out- of-range areas can be filled with a solid color. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 56
Times at A/B cursors The following items appear on the [Item settings] tab. • Display on/off • Graph line colors and thickness • Bar graph display on/off [Item settings] HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 57
Scrolls to display the left edge of the table. Scrolls to the right edge of the table. Copies the value of the currently selected cell to the clip- Press Ctrl keys simultaneously board. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 58
Select all items (up to 600) of the same property. Sort by model name, serial no., or model comment. Sort Items Move blue mouse-selected items up or down. Restore original order. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
The currently selected logger’s back- ground is a different color. If multiple devices are listed, they can Returns to the be sorted in ascending order. main screen. p.57 See: HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 60
The Save Method screen will appear when the [Always specify folder and file before importing] checkbox on the optional screen is selected. "7.1 Changing the Saving Method for Imported Data" (p.78) See: HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 61
(p.54) Click the button to display the table. The Logger Settings screen appears. Returns to the main screen. Displays the Data Import screen (p.55). HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Select the items to be displayed in the graph (p.54). Information about the latest data Click to select. The currently selected logger’s background is a different color. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 63
Recently folder Click The last ten folders containing data that was displayed as a graph or ta- ble are listed. Select the drive Select the file Select the folder HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
How can I print only part of a graph? Click Specify the time span to print, and click [Print] button. Times that are not displayed are not printed. "Viewer Screen" (p.47) See: Example Graph Printout HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Select the items for processing (p.54). Information about the latest data Click to select. The currently selected logger's background is a different color. Continued → HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 66
Performs approximate two-data-item arithmetic calculation. (p.69) Arithmetic OVER Data Converts data outside of the upper and lower threshold set- (p.70) tings to specified values, and saves as new data items. Revision HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
[Calculate] button to display the scaled result. confirmation 3. Click the [Execute] button. (The scaled results are saved.) Note: Click the [Finish] button to close the [Scaling] dialog box. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
•To delete a setting, click it then click the [Delete] button. 3. Click the [Execute] button. (Calculation results are saved.) Note: Click the [Finish] button to close the [Power Calculation] dialog box. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
2. Specify the cost per kWh, voltage, and power factor. 3. Click the [Calculate] button. (Electric power consumption and energy cost values are calculated and displayed.) Note: Click the [Finish] button to close the [Energy Cost] dialog box. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
2. Set the upper threshold. 3. Click the [Calculate] button. (Operating hours and operating rate values are calculated and displayed.) Note: Click the button to close the [Operating Rate] dialog box. [Finish] HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
1. Select the item and time span. 2. Click the [Execute] button. (Integration results are saved.) Note: Click the [Finish] button to close the [Integration] dialog box. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
1. Select the items and time span. 2. Click the [Execute] button. (Calculation results are saved.) Note: Click the [Finish] button to close the [Dew Point] dialog box. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
1. Select the items and time span. 2. Select the calculation operator. 3. Click the [Execute] button. (Calculation results are saved.) Note: Click the [Finish] button to close the [Two-Data-Item Arithmetic] dialog box. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
2. Set the upper and lower threshold values, and their corresponding conversion val- ues. 3. Click the [Execute] button. (Conversion results are saved.) Note: Click the button to close the dialog box. [Finish] [OVER Data Revision] HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
The last ten folders containing data that was displayed as a graph or table are listed. Select the desired file. Returns to the Select main screen. Clears all selections (removes check marks). HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
6.1 Copying and Moving Data 6.1 Copying and Moving Data The selected logger recording files can be copied or moved to any folder. Example: Copy a file from the folder C:\Users\hioki\Documents\LR5000 to C:\Users\hioki\Desktop. Select [Copying Data] [Moving Data]. Click Select the drive.
6.2 Deleting Data 6.2 Deleting Data Select and delete logger recording files as follows. Example: Delete a file from the folder C:\Users\hioki\Documents\LR5000. Select Click [Deleting Data]. Select the drive. Select the file. (Up to 100 can be selected.) Select the folder.
6.3 Combining Data 6.3 Combining Data Separate logger recording files can be combined into one set of recording data. Example: Combine file 20110117 with other files in C:\Users\hioki\Docu- ments/LR5000, and save the combined data file in C:\Users\hioki/ Desktop. Select [Combining Click Data].
Click Select [Extracting Data]. Select the drive. Select the file. (one only) Select the folder. Specify the extracting time span and extracted data (model). HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
[Automatically import and store data when the logger is connected to a computer] checkbox and then clear and the [Always specify folder and file before importing] check box to display the Data Import screen ( p.56). HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
File names can be specified as a com- bination of up to three of these compo- nents: model name, model comment, serial no., re- cording start day, recording start date and time, and save day HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
COM port to monitor. How can the function settings of the logger's settings displays be changed? [Save Settings] Show the settings of the [Open Settings] buttons. ( p.38) HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
• Temperature: 10 to 50°C (14.0 to 122.0°F) (logger, sensor) • Humidity: 80%RH or less (logger), 10 to 90%RH (sensor), non-con- densating Product warranty period 3 years Guaranteed accuracy period 1 year HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at:...
Data Collector The measurement data display turns off about 30 seconds after the last Power save setting button operation (cancel power save for continuous display) Real-time clock Provided HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
158.0°F), 80%RH or less (non-condensating) • Storage temperature and humidity: -20°C to 70°C (-4.0°F to 158.0°F), 80%RH or less (non-condensating) • Safety: EN61010 Applicable Standards • EMC : EN61326 HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Converts between the logger's infrared signals and USB signals to sup- Functions port communications between the logger and a computer (USB port). LR5001 Humidity Logger, LR5011 Temperature Logger, LR5031 Instru- mentation Logger, LR5041 Voltage Logger (50 mV), LR5042 Voltage Logger (5 V), LR5043 Voltage Logger (50 V), LR5051 Clamp Logger...
Page 87
A small resident program (icon in the task tray/notification area) detects Auto-start function when a logger or the Data Collector is connected to the computer, and automatically starts the LR5000 Utility Program. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at:...
Page 88
(the following functions are supported by the supplied PC Utility version File management 2.00, or later) functions • Delete data saved to an SD memory card in the LR5092-20 Data Col- lector Help function Displays helpful operating instructions HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Storage temperature and humidity • Humidity: 80%RH or less (non-condensating) (Connects to the LR5001 Humidity Logger) • Temperature: between -40.0 and 0.0°C (-40.0 to 32.0°F): ±1.0°C (±1.8°F) between 0.0 (32.0°F) and 35.0°C (95.0°F): ±0.5°C (±0.9°F) between 35.0 (95.0°F) and 70.0°C (158.0°F): ±1.0°C (±1.8°F) between 70.0 (158.0°F) and 85.0°C (185.0°F): ±2.0°C (±3.6°F)
Page 90
°C solvents or corrosive gases, or an environment of high temperature/humidity (above 50 60% RH). HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at:...
We do not take any responsibility for damage incurred during shipping. • Please contact your dealer or Hioki representative for information on where to sub- mit products for repair. When the logger will not be used for long time...
9.2 Disposing of the Logger Obey local regulations for disposal of electronic equipment. 9.3 Troubleshooting If damage is suspected, check the "Before requesting repairs" section before con- tacting your dealer or Hioki representative. Before requesting repairs Problem Symptom Probable Causes...
Page 94
The maximum and minimum val- Please contact your dealer or Hioki ues are not displayed when the representative. recorded data count is 0. See: "Requesting repairs"...
Page 95
84000 data sets per measure- tially.] graphs and tables. ment parameter. Change the duration to be displayed. Change from [All Data] to [1day]. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
The display appears as follows when an error occurs on the logger. Logger Error Displays Error Displays Meaning Remedies and References Calibration data error: Inspection and repair is required. Please contact your dealer or Hioki A fault occurred with the internal representative. calibration data. See: "Requesting repairs" (p.91) Microcomputer error: A fault occurred in microcomputer ROM/RAM.
Instantaneous values Maximum, minimum, and average values over are recorded this time span are recorded. Statistical recording cannot be selected when the recording interval is set to one second. HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
20 sec. 3 d, 11 h, 20 min 20 min 208 d, 8 h 30 sec. 5 d, 5 h 30 min 312 d, 12 h 60 min 625 d HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Recording Interval Battery Life Recording Interval Battery Life 1 sec. to 30 sec. Approx. 20 days 1 min Approx. 3 months 10 min or more Approx. 1 year HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
......... 17 Data import screen ........55 Installing the logger ........40 DATA indicator .......... 13 Installing the PC application program ..21 Delete ............73 Instantaneous recording ....35 HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 101
(for when memory becomes full) .... 30 Part names/functions ........12 TIME indicator ........13 PC application program Time setting ......... 15 Installation ..........21 Troubleshooting ........92 Operating requirements ......21 Screens ..........21 Uninstall ..........23 HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 102
View latest data ........58 Viewer ....... 46 Wall-mounted holder ........ 41 When the logger will not be used for long time ..........91 Year, month, day, and hour setting ..15 HIOKI LR5001B980-05 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...