Wireless Module Specifications
Measurement
target
Thermocouple
(Not including
accuracy of
reference
junction
compensation)
Find Quality Products Online at:
Maximum
Type
Range
resolution
E
100°C f.s.
0.01°C
500°C f.s.
0.05°C
2000°C f.s.
0.1°C
T
100°C f.s.
0.01°C
500°C f.s.
0.05°C
2000°C f.s.
0.1°C
GlobalTestSupply
www.
Measurable range
Greater than or equal to
−100°C and less than 0°C
Greater than or equal to 0°C
and less than or equal to
100°C
Greater than or equal to
−200°C and less than
−100°C
Greater than or equal to
−100°C and less than 0°C
Greater than or equal to 0°C
and less than or equal to
500°C
Greater than or equal to
−200°C and less than
−100°C
Greater than or equal to
−100°C and less than 0°C
Greater than or equal to 0°C
and less than or equal to
1000°C
Greater than or equal to
−100°C and less than 0°C
Greater than or equal to 0°C
and less than or equal to
100°C
Greater than or equal to
−200°C and less than
−100°C
Greater than or equal to
−100°C and less than 0°C
Greater than or equal to 0°C
and less than or equal to
400°C
Greater than or equal to
−200°C and less than
−100°C
Greater than or equal to
−100°C and less than 0°C
Greater than or equal to 0°C
and less than or equal to
400°C
.com
sales@GlobalTestSupply.com
Measurement
accuracy
±0.7°C
±0.5°C
±0.9°C
±0.7°C
±0.5°C
±0.9°C
±0.7°C
±0.5°C
±0.7°C
±0.5°C
±1.4°C
±0.7°C
±0.5°C
±1.4°C
±0.7°C
±0.5°C
Need help?