Configuring and Establishing a USB Connection
Connecting the instrument to the computer with a USB cable
This section describes how to connect the instrument to a computer with a USB cable.
Instructions
You cannot use both the Logger Utility and communication commands at the same time over
the USB connection. Allow at least 30 s to switch between the Logger Utility and communication
commands.
Find Quality Products Online at:
Do not unplug the USB cable while the instrument is sending or receiving data.
Doing so could damage the instrument.
Turn off the instrument and computer before connecting or disconnecting the
USB cable.
Failure to do so could cause the instrument to malfunction or damage it.
Ensure the instrument and computer are grounded at the same potential.
Connecting the USB cable while the ground potentials differ between the instrument
and computer could cause the instrument to malfunction or damage it.
GlobalTestSupply
www.
NOTICE
1
Insert the USB cable plug into the
instrument's USB cable connector,
exercising care to ensure that the plug is
oriented properly.
2
Connect the other end of the USB cable
to the computer's USB connector.