Measurement
Thermocouple
(Not including
accuracy of
reference
junction
compensation)
Find Quality Products Online at:
Type
Range
target
E
100°C f.s.
500°C f.s.
2000°C f.s.
T
100°C f.s.
500°C f.s.
2000°C f.s.
N
100°C f.s.
500°C f.s.
2000°C f.s.
GlobalTestSupply
www.
Maximum
Measurable range
resolution
0.01°C
Greater than or equal to −100°C
and less than 0°C
Greater than or equal to 0°C and
less than or equal to 100°C
0.05°C
Greater than or equal to −200°C
and less than −100°C
Greater than or equal to −100°C
and less than 0°C
Greater than or equal to 0°C and
less than or equal to 500°C
0.1°C
Greater than or equal to −200°C
and less than −100°C
Greater than or equal to −100°C
and less than 0°C
Greater than or equal to 0°C and
less than or equal to 1000°C
0.01°C
Greater than or equal to −100°C
and less than 0°C
Greater than or equal to 0°C and
less than or equal to 100°C
0.05°C
Greater than or equal to −200°C
and less than −100°C
Greater than or equal to −100°C
and less than 0°C
Greater than or equal to 0°C and
less than or equal to 400°C
0.1°C
Greater than or equal to −200°C
and less than −100°C
Greater than or equal to −100°C
and less than 0°C
Greater than or equal to 0°C and
less than or equal to 400°C
0.01°C
Greater than or equal to −100°C
and less than 0°C
Greater than or equal to 0°C and
less than or equal to 100°C
0.05°C
Greater than or equal to −200°C
and less than −100°C
Greater than or equal to −100°C
and less than 0°C
Greater than or equal to 0°C and
less than or equal to 500°C
0.1°C
Greater than or equal to −200°C
and less than −100°C
Greater than or equal to −100°C
and less than 0°C
Greater than or equal to 0°C and
less than or equal to 1300°C
.com
Measurement
accuracy
±0.7°C
±0.5°C
±0.9°C
±0.7°C
±0.5°C
±0.9°C
±0.7°C
±0.5°C
±0.7°C
±0.5°C
±1.4°C
±0.7°C
±0.5°C
±1.4°C
±0.7°C
±0.5°C
±1.1°C
±0.9°C
±2.1°C
±1.1°C
±0.9°C
±2.1°C
±1.1°C
±0.9°C
sales@GlobalTestSupply.com
313
Need help?
Do you have a question about the MEMORY HiLOGGER LR8450 and is the answer not in the manual?