Measurement
Thermocouple
(Not including
accuracy of
reference
junction
compensation)
Humidity
Resistance
temperature
detector
Resistance
Reference junction
compensation
Temperature
characteristics
Find Quality Products Online at:
Type
Range
target
B
2000°C f.s.
C
100°C f.s.
500°C f.s.
2000°C f.s.
–
100% RH f.s. 0.1% RH
Humidity accuracy table
Each boundary line is included in the better measurement accuracy area.
Pt100
100°C f.s.
500°C f.s.
2000°C f.s.
JPt100
100°C f.s.
500°C f.s.
2000°C f.s.
Pt1000
100°C f.s.
500°C f.s.
2000°C f.s.
Ω
–
10
Ω
20
100
200
±0.5°C (with input terminal temperature balancing)
Reference junction compensation: Add thermocouple measurement accuracy when set to
internal.
Add (measurement accuracy × 0.1) per degree to the measurement accuracy figure (for
information about the humidity accuracy, see the humidity accuracy table).
GlobalTestSupply
www.
Maximum
resolution
0.1°C
Greater than or equal to
400°C and less than 600°C
Greater than or equal to
600°C and less than 1000°C
Greater than or equal to
1000°C and less than or
equal to 1800°C
0.01°C
0°C to 100°C
0.05°C
0°C to 500°C
0.1°C
0°C to 2000°C
5.0% RH to 95.0% RH
100
95
±10% RH
80
±8% RH
60
±6% RH
5
0
0
: Out of accuracy guarantee range
0.01°C
−100°C to 100°C
0.05°C
−200°C to 500°C
0.1°C
−200°C to 800°C
0.01°C
−100°C to 100°C
0.05°C
−200°C to 500°C
0.1°C
−200°C to 500°C
0.01°C
−100°C to 100°C
0.05°C
−200°C to 500°C
0.1°C
−200°C to 800°C
Ω
Ω
f.s.
0.5 m
0
Ω
Ω
f.s.
1 m
0
Ω
Ω
Ω
f.s.
5 m
0
Ω
Ω
Ω
f.s.
10 m
0
Measurement
Measurable range
±5.4°C
±3.7°C
±2.4°C
±1.7°C
±1.7°C
±1.7°C
As per
measurement
accuracy of the
Z2000 Humidity
Sensor
±8% RH
±10% RH
±6% RH
±8% RH
±5% RH
±6% RH
20
30
Temperature (°C)
±0.5°C
±0.7°C
±0.9°C
±0.5°C
±0.7°C
±0.9°C
±0.5°C
±0.7°C
±0.9°C
Ω
to 10
±10 m
Ω
to 20
±20 m
Ω
to 100
±100 m
Ω
to 200
±200 m
.com