Data Handling
11.15 Data Handling
In the following circumstances, calculated values and saved data are treated as shown in the table
below:
• When the waveform significantly exceeds the range's measurable range (+OVER, -OVER)
• When communications are temporarily interrupted (NO DATA)
• When the instrument detects a thermocouple wire break during temperature measurement (wire
break detection)
Input type
Voltage
Thermocouple
Resistance
temperature
detector
Humidity
Resistance
Strain
Integration
Find Quality Products Online at:
Input range
+OVER
1 mV
0.00163835
2 mV
0.0032767
5 mV
0.00819175
10 mV
0.0163835
20 mV
0.032767
50 mV
0.0819175
100 mV
0.163835
200 mV
0.32767
1 V
1.63835
2 V
3.2767
10 V
16.3835
20 V
32.767
100 V
163.835
1-5 V
16.3835
100°C
327.67
500°C
1638.35
2000°C
3276.7
100°C
327.67
500°C
1638.35
2000°C
3276.7
100% RH
3276.7
Ω
10
16.3835
Ω
20
32.767
Ω
100
163.835
Ω
200
327.67
1000 με
1638.35
2000 με
3276.7
5000 με
8191.75
10000 με
16383.5
20000 με
32767
50000 με
81917.5
100000 με
163835
200000 με
327670
1000 Mc
2147483647
GlobalTestSupply
www.
−OVER
NO DATA
- 0.0016384
- 0.0032768
- 0.008192
- 0.016384
- 0.032768
- 0.08192
- 0.16384
- 0.32768
- 1.6384
- 3.2768
- 16.384
16.3825
- 32.768
32.765
- 163.84
163.825
- 16.384
16.3825
- 327.68
327.65
- 1638.4
1638.25
- 3276.8
3276.5
- 327.68
327.65
- 1638.4
1638.25
- 3276.8
3276.5
- 3276.8
3276.5
- 16.384
16.3825
- 32.768
32.765
- 163.84
163.825
- 327.68
327.65
- 1638.4
1638.25
- 3276.8
3276.5
- 8192
8191.25
−16384
16382.5
−32768
32765
−81920
81912.5
−163840
163825
−327680
327650
–
.com
Wire break
detection
0.00163825
–
0.0032765
–
0.00819125
–
0.0163825
–
0.032765
–
0.0819125
–
0.163825
–
0.32765
–
1.63825
–
3.2765
–
–
–
–
–
327.66
1638.3
3276.6
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
sales@GlobalTestSupply.com