Hioki MEMORY HiLOGGER LR8450 Instruction Manual page 336

Hide thumbs Also See for MEMORY HiLOGGER LR8450:
Table of Contents

Advertisement

Digital filters
Power supply
frequency
filter setting
60 Hz
50 Hz
-2. Accuracy specifications
Accuracy guarantee
conditions
Measurement range, maximum resolution, measurable range, and measurement accuracy figures
Measurement
Voltage
Temperature
characteristics
Normal-mode
rejection ratio
Common-mode
rejection ratio
Find Quality Products Online at:
The digital filter cutoff frequency is automatically set as described in the following table
according to the data refresh interval and power supply frequency filter setting:
1 ms 2 ms 5 ms
10 ms 20 ms 50 ms 100 ms 200 ms 500 ms
50 k
5.4 k 2.6 k
1.0 k
50 k
5.4 k 2.6 k
1.0 k
Accuracy guarantee
duration
Accuracy guarantee
duration after
adjustment made by
Hioki
Accuracy guarantee
temperature and
humidity range
Warm-up time
The conditions have been specified with zero-adjustment finished and cutoff frequency
set to 5 Hz, 10 Hz, 50 Hz, or 60 Hz (see cutoff frequency table in "Digital filters" [p.  3 31]).
Type
Range
target
100 mV f.s.
200 mV f.s.
1 V f.s.
2 V f.s.
10 V f.s.
20 V f.s.
100 V f.s.
1-5 V f.s.
Add (measurement accuracy × 0.1) per degree to the measurement accuracy figure.
50 dB or greater
(With 50 Hz input, power supply frequency filter of 50 Hz, and a data refresh
interval of 5 s)
(With 60 Hz input, power supply frequency filter of 60 Hz, and a data refresh
interval of 5 s)
Signal source resistance of 100
100 dB or greater
(With 50 Hz/60 Hz input and a data refresh interval of 1 ms)
140 dB or greater
(With 50 Hz input, power supply frequency filter of 50 Hz, and a data refresh
interval of 5 s in the 100 mV f.s. range)
(With 60 Hz input, power supply frequency filter of 60 Hz, and a data refresh
interval of 5 s in the 100 mV f.s. range)
GlobalTestSupply
www.
Data refresh interval
400
200
100
60
400
200
100
50
1 year
1 year
23°C ±5°C (73°F ±9°F), 80% RH or less
At least 30 min. after connecting to the LR8450/LR8450-01
Memory HiLogger and turning on the instrument
Maximum
Measurable range
resolution
5 μV
−100 mV to 100 mV
10 μV
−200 mV to 200 mV
50 μV
−1 V to 1 V
100 μV
−2 V to 2 V
500 μV
−10 V to 10 V
1 mV
−20 V to 20 V
5 mV
−100 V to 100 V
500 μV
1 V to 5 V
Ω
or less
.com
1 s
2 s
5 s 10 s
60
10
5
5
5
50
10
5
5
5
Unit: Hz
Measurement
accuracy
±100 μV
±200 μV
±1 mV
±2 mV
±10 mV
±20 mV
±100 mV
±10 mV
sales@GlobalTestSupply.com
331

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MEMORY HiLOGGER LR8450 and is the answer not in the manual?

This manual is also suitable for:

Memory hilogger lr8450-01Lr8450Lr8450-01

Table of Contents