Window triggers
This section describes how to specify a range (window) using upper and lower limit values and then
activate a trigger when the waveform moves into or out of that range.
You can activate a trigger either when the waveform enters the range (window IN) or when the
waveform exits the range (window OUT).
>
1
Under
A window trigger will be used.
2
Choose between
IN
OUT
When
range.
3
Under
The range defined by the upper and lower limit values will serve as the window.
When using the scaling function, post-scaling values are displayed.
Find Quality Products Online at:
Window IN
> [Unit
n],
[Remote n]
1 1
3 3
1 1
[Type]
for
[Start]
or [Stop], set the trigger type to [Window].
[IN]
and
[OUT]
Activates the trigger when the waveform enters the specified range.
Activates the trigger when the waveform exits the specified range.
[Condition]
is set to [AND], the system will determine whether the waveform is inside the specified
[Upper]
and [Lower], set the upper and lower limit values, respectively.
GlobalTestSupply
www.
Analog Triggers, Pulse Triggers, Waveform Calculation Triggers
Window OUT
(n = 1, 2, . . .)
2
2
for the waveform.