5
Under [Format], select the file format.
Binary
Text
MDF(MF4)
If
[Text]
See
[Settings...]
6
Under [Range], select the range of data to save.
All
A-B
Start-A
Start-B
A-End
B-End
See "Specifying a waveform range" (p. 9 7).
(When
Under [Downsampling], select whether to perform downsampling.
When set to [ON], the size of the saved data file will be reduced.
OFF
(When
Under
1/2
Example: If set to [1/5], 1 data point out of every 5 data points will be left.
(When
Under
Instant
Statistics
7
Press the
Waveform data will be saved.
Find Quality Products Online at:
Saves data in the instrument's dedicated format (binary format).
You will be able to load binary-form data on the instrument and with the Logger Utility.
Saves data in text format.
Although you will be able to load text-form data with spreadsheet software, you will not
be able to load it on the instrument or with the Logger Utility.
Saves waveform data in the MDF format (Ver. 4).
These files can be loaded on a PC using a waveform viewer that supports the MDF
format. They cannot be loaded by the instrument or the Logger Utility.
is selected, configure the text format settings while
of
[Text]
in "Auto save (real-time save)" (p. 1 46).
Saves the waveform for the entire recorded length, without regard to the A/B cursors.
Saves the waveform between the A and B cursors.
Saves the waveform from the beginning to the A cursor.
Saves the waveform from the beginning to the B cursor.
Saves the waveform from the A cursor to the end.
Saves the waveform from the B cursor to the end.
[Format]
is set to [Text])
,
ON
[Downsampling]
is set to [ON])
[Decimation
factor], select the decimation factor.
1/100000
to
[Downsampling]
is set to [ON])
[Save
data], select the data thinning method.
Saves the first data point.
Example: With a setting of [1/5], only the first of each group of five data points will be
saved.
Saves statistical data (maximum value, minimum value, and average value).
Example: With a setting of [1/5], the maximum, minimum, and average values will be
saved for each set of five data points.
ENTER
key while
[Save]
GlobalTestSupply
www.
[Settings...]
is selected.
.com
Saving Data
is selected.
sales@GlobalTestSupply.com