2.2.2 Operation Panel

Name
Stand-by Switch
Status Display LED
Cassette Ready
Lamp
Cassette Removal
Lamp
Excessive amount of
X-ray Display Lamp
Power/Erasing Mode
Indicator Lamp
Eraser Button
Turns on/off the power to the FCR PRIMA T2 (CR-IR 392). This switch
is used when the FCR PRIMA T2 (CR-IR 392) is started up singly.
When the buzzer sounds because of an error or detection of excessive
amount of X-rays and this switch is pressed, the buzzer stops and the
cassette lock is released.
To turn off the unit, keep pressing the switch for about four seconds.
Indicates status of the unit.
: Initializing
: Console unconnected
Blinking
: Error/Warning indication
number
For details, see [4.4 When An Error Code Appears] in
the Reference Guide.
Lights up (green) when it is ready for inserting a cassette. You can not
insert the cassette when this lamp is off.
When the lamp is on alternately with the Cassette Removal Lamp, an
unprocessed cassette is ejected.
Blinks (blue) when it is ready for ejecting the processed cassette. Lights
off when you remove the cassette. When this lamp is on alternately with
the Cassette Ready Lamp, an unprocessed cassette is ejected.
When this lamp is on alternately with the Excessive amount of X-ray
Display Lamp, the Excessive amount of X-ray erasing-incomplete
cassette is ejected. Press the Eraser Button and Perform an erasion.
Note that it takes a lot of time to achieve erasure.
If X-ray of about 400 mR and over is detected, this lamp blinks
alternately with the Cassette Removal Lamp, and the excessive amount
of X-ray erasing-incomplete cassette is ejected. Press the Eraser Button
and perform an erasion. Note that it takes a lot of time to achieve erasure.
If erasing is to be done later, remove the cassette. If the cassette is
to be used the next time, always perform erasing after 16 hours have
passed before using the cassette.
Lights up with green color when the main power switch is set to ON and
turns to orange color under the erasing mode.
Switches the mode from reading to erasing.
When you erase IP images, press this button to switch to the erasing mode.
Function
: End processing
: Console connected
FCR PRIMA T2 Operation Manual