Making Measurements
Setting Up the Trigger and Making Measurements
Setting Up the Trigger and Making Measurements
Trigger Model
Figure 5-1 shows the trigger model used in the E5070/E5071A.
Figure 5-1
Trigger Model Used in the E5070/E5071A
The E5070/E5071A has one trigger source. The trigger signal generated by this trigger
source will be sent to channels 1 through 9 in order. The trigger mode is set independently
for each channel. If the trigger mode is set to "Hold," sending a trigger will not prompt a
sweep.
NOTE
Executing measurement of a channel does not depend on the channel's display status. You
can even perform measurement of a channel with no display.
The traces actually measured are set by the number of traces in each channel.
122
Chapter 5