Handler Interface Functional Test

The 4349B's handler interface function is tested using the built-in selftest and the handler
interface tester.

Test Equipment

Description
Handler Interface Tester

Procedure

Initial Setup.

1. Turn the 4349B OFF.
2. Set the LED PW switch on the Handler Interface Tester to OFF , and set the IN1, IN2, IN3,
and IN4 switches to 0 .
3. Set up the equipment as shown in Figure 9-10
4. Turn the 4349B ON.
5. Reset the 4349B using the following procedure.
a. Press
b. Select
using

Key Lock Function Test.

6. Set the IN4 switch on the Handler Interface Tester to 1 .
7. Con rm that the all keys on the 4349B's front panel are locked out.
8. Set the IN4 switch on the Handler Interface Tester to 0 .

External Trigger Function Test.

9. Press
to set the trigger mode to External.
Recommended Model
04339-65007
Figure 9-10. Handler Interface Functional Test Setup
to display the system reset settings.
or
and press
.
Maintenance
9-23