If Gain Adjustment; Receiver Characterization - Keysight Technologies N5224/5A Service Manual

2-port and 4-port pna - microwave network analyzers
Table of Contents

Advertisement

Tests and Adjustments
Adjustments
Procedure
1. Connect the equipment as shown in
between the network analyzer and the power meter.
Figure 3-15
Equipment Setup for the Source Adjustment
2. .Press UTILITY
Source Adjustment.
3. Ensure the GPIB settings are correct.
4. Follow the instructions and prompts as they are displayed.

IF Gain Adjustment

The IF gain adjustment is used to adjust the IF gain of the network analyzer.
Procedure
1. Press UTILITY
Gain Adjustment.
2. Follow the instructions and prompts as they are displayed.

Receiver Characterization

This characterizes the receivers in your analyzer.
Procedure
1. Press UTILITY

Receiver Characterization.

Keysight N5224A/25A Service Guide
Figure
Service
System
, then
Service
System
, then
Service
System
, then
3-15. Connect a GPIB cable
Adjustments
, then
, then click
Adjustments
, then
, then click IF
Adjustments
, then
, then click
3-43

Advertisement

Table of Contents
loading

Table of Contents