Supplement for the User's Guide (Firmware Version 7.66)
Conversion Loss Using the Frequency Offset Mode
Performing a Power Meter Calibration Over the RF Range
1. Make the connections as shown in
meter calibration over the RF range.
Figure 1-20
8753ET/ES Power Meter Calibration Equipment Setup
for the RF Range
Figure 1-21
8753ES Option 011 Power Meter Calibration Equipment Setup
for the RF Range
2. Use the power meter settings that were set up in steps 1 through 4 of the
Power Meter (Source) Calibration Over the IF Range"
Perform a one-sweep power meter calibration over the RF frequency range at 0 dBm by
pressing
PWRMTR CAL
Cal
NOTE
Do not reduce the number of points to perform this power meter calibration.
Reducing the number of points will turn off the receiver calibration.
1-28
Figure 1-20
(or
Figure
ONE SWEEP
0
x1
1-19) for the one-sweep power
"Performing a
section.
TAKE CAL SWEEP
.