Calibrating for Increased Measurement Accuracy
Adapter Removal Using ECal (ES Analyzers Only)

Perform the 2-Port Error Corrections

1. Connect adapter A3 to adapter A2 on port 2 as shown in
Figure 6-27 Two-Port Cal Set 1
2. Connect the ECal module between adapter A1 and adapter A3.
3. Press
Cal
ECal MENU
4. Press
FULL 2-PORT
NOTE
When using adapter removal calibration, you must save calibration sets to the internal disk,
not to internal memory.
5. Save the results to disk. Name the file "PORT1."
6. Remove the ECal module and adapter A3 from the setup. Connect adapter A3 to adapter A1 on port 1 as
shown in
Figure
6-28.
6-70
.
MODULE A b
to perform the first 2-port error correction using the ECal module.
Figure
6-27.