Calibrating for Increased Measurement Accuracy
Frequency Response Error Corrections

Response Error Correction for Transmission Measurements

1. Press
.
Preset
2. Select the type of measurement you want to make.
If you want to make a transmission measurement in the forward direction (S
press:
Trans: FWD S21 (B/R)
Meas
For ES analyzers, if you want to make a transmission measurement in the reverse
direction (S
), press:
12
Trans: REV S12 (A/R)
Meas
3. Set any other measurement parameters that you want for the device measurement:
power, number of points, IF bandwidth.
4. To select a response correction, press:
CALIBRATE MENU
Cal
5. Make a "thru" connection between the points where you will connect your DUT.
NOTE
Include any adapters or cables that you will have in the device measurement.
That is, connect the standard device where you will connect your DUT.
Figure 6-3 Standard Connections for Response Error Correction for
Transmission Measurements
6. To measure the standard, press:
THRU
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard
measurement. The analyzer underlines the
calibration standard, and computes the error coefficients.
6-14
or on ET models:
RESPONSE
THRU
TRANSMISSN
softkey after it measures the
),