Thermodisinfecting The Suction Hoses - Sirona C2+ Cleaning And Care

Hide thumbs Also See for C2+:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Care and cleaning instructions for the practice team C2
3.6

Thermodisinfecting the suction hoses

Grease
59 58 512 D 3370
D 3370.103.01.13.02
+
+
, M1
The unit is usually equipped with suction hoses that can-
not be thermodisinfected.
However, it may have suction hoses that can be ther-
modisinfected as special accessories. These are
Grease
marked in turquoise color (10).
ATTENTION
Only hoses marked with a turquoise ring (10) can be
thermodisinfected!
Switch the treatment center OFF.
10
Pull off the suction hoses and the connection hose
from the assistant element and the water unit.
Remove the top parts of the filter housings 5 and the
amalgam residue collector 6.
Pull off the handpiece (7) from the suction hose.
Grease
For the saliva ejector hose, insert the intermediate piece
(8) in the hose holder (9).
Lubricate the O-rings.
ATTENTION
Use only lubricants which have been approved by Sirona
(see page 2, General information)!
Fit the hose holders onto the rails in the thermodisinfec-
tor (by Miele) as illustrated.
Connect the suction hoses to the hose holders and place
them in the wire basket (by Miele) intended for thermo-
disinfection.
Thermodisinfect at max. 93×C (197°F).
Before refitting the hoses, grease the O-rings on the
connection pieces.
The hoses must engage firmly.
Hose holder for Miele thermodisinfector
Order No. 89 18 757
3 Vacuum system
3.6 Thermodisinfecting the suction hoses
11