Instructions for Use

Stopping the TPS-mobile

You can switch the pumps off by using the "pumping" selector (moving it to
position "0") – the TMP controller will remain ON and if a gauge is connected to
the system it will be possible to continue reading the pressure even if pumps
are off.
An alternative mode for switching the pumps off is the use of Main Switch
(move it to position "0") – The system will be completely switched off and no
reading or operation will be possible.
Before carrying out any work on the system, disconnect it from the power
WARNING
supply, vent the pump with the vent valve, wait until the rotor is completely
stopped and wait until the surface temperature of the pump is below 50° C.

Emergency stop

To immediately stop the TPS-mobile in an emergency condition it is necessary to
remove the supply cable from the mains plug or moving the external switch to OFF
position.
This mode is to be used only in an emergency. The use of this stop mode
WARNING
could lead to faults and/or damages at the pump. Improper use of this
stop mode could void the warranty.

Modified Standard

TPS-mobile platform is suited to be tailored in according to many different needs so
special TPS-mobile PNs can be released.
Agilent provides support for many different configurations, descriptions of which
can be found in the Orderable Parts section of this manual. Nevertheless, if you
have difficulty finding a P/N that matches your needs, please contact your Sales
Representative.