Running Diagnostics
Overview of Diagnostics Testplans
Configuring the system.ust file in the System Configuration
Editor
N O T E
N O T E
20
1 Start the System Configuration Editor by clicking this
icon on your desktop:
2 Click on the detected on system button. The screen will
show instruments detected on the system. Any instrument
in
Table 3- 1
which is in your system should be detected.
3 Double- click on the instrument name to add it to the
system.ust file. It is recommended that you add
instruments in the following order:
i the SLU,
ii the first pin matrix card in the SLU, and
iii other instruments in any order.
At least 1 SLU, 1 Pin Matrix, and 1 DMM must be in the system and in the
.ust file for the DGN and CEDGN testplans to run.
4 Once the instruments have been added, click Save or
Save As if you would like to change the name of your .ust
file.
Ensure that you have added all the standard instruments you would like to
test. The DGN and CEDGN testplans will only test instruments in the .ust
file.
TS-8989 Diagnostics User's Guide