13:00 12-15-05 flags C000000 conc 0.000 tgflo 30.000
agflo 0.000 tzflo 1000.000 azflo 0.000 ttflo 1030.000
atflo 0.000 pres 0.000
14:00 12-15-05 flags C000000 conc 0.000 tgflo 30.000
agflo 0.000 tzflo 1000.000 azflo 0.000 ttflo 1030.000
atflo 0.000 pres 0.000
erec format
lrec format
srec format
These commands report the output format for long and short records, and
dynamic data in various formats such as ASCII without text, ASCII with
text, or binary. The example below shows the output format for long
records is ASCII with text, according to
Send:
lrec format
Receive:
lrec format 01
set erec format format
set lrec format format
set srec format format
These commands set the output format for long and short records, and
dynamic data, according to
record output format to ASCII with text.
Send:
set lrec format 1
Receive:
set lrec format 1 ok
Table B–4. Record Output Formats
Format
Output Format
0
ASCII no text
1
ASCII with text
2
Binary data
erec layout
lrec layout
srec layout
These commands report the layout (string indicating the data formats) for
data that is sent out in response to the erec, lrec, srec, and related
commands. For details on how to interpret the strings, see "Record Layout
Definition" later in this appendix.
Table
B-4.
Table
B-4. The example below sets the long
Model 146i Instruction Manual
C-Link Protocol Commands
Datalogging