Download Print this page

Figure 7-14. Removing The Glass Chamber Assembly - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

Thermo Fisher Scientific
g. Tighten knurled screw finger tight. Do not use tools to tighten.
h. Replace oven cover, being careful to place tubing and wire in slot of
cover.
6. Install the new oven by following the previous steps in reverse.
7. Check the system for leaks. Refer to "Leak Test and Pump Checkout" in
this chapter.
Top Assembly
Glass
Chamber

Figure 7-14. Removing the Glass Chamber Assembly

Permeation Oven Replacement
Knurled Screw
Glass Chamber
Assembly
Oven Top
Oven Enclosure
Model 146i Instruction Manual
Servicing
7-47

Advertisement

loading