Table B-8. Standard Ranges - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

Thermo Fisher Scientific
This command sets the ozonator target concentration under photometer
control, in ppb. This command is only available if the photometer option is
installed.
Send:
set o3 conc 500
Receive:
set o3 conc 500 ok
o3 setting
This command reports the ozonator target concentration under
photometer control, in ppb. The example below reports a setting of 100
ppb. This command is only available if the photometer option is installed.
Send:
o3 setting
Receive:
o3 setting 1.000E+02 ppb
range
This commands report photometer O
instrument responds with "can't, wrong settings". The example below
reports that the O
full-scale range is set to 5.0 ppm, according to
3
8. This command is only available if the photometer option is installed.
Send:
range
Receive:
range 6: 500E+03 ppb
set range selection
This command selects the O
The example below sets the O
Send:
set range 5
Receive:
set range 5 ok
Table B–8. Standard Ranges
Selection
ppm
0
0.05
1
0.1
2
0.2
3
0.5
4
1.0
5
2.0
6
5.0
7
C1
8
C2
9
C3
range. If the mode is incorrect, the
3
full-scale ranges, according to
3
full-scale range to 2.0 ppm.
3
Model 146i Instruction Manual
C-Link Protocol Commands
Measurement Configuration
Table B–
Table
B–8.
B-35