Measurements

Measurements
B-8
Model 146i Instruction Manual
Command
Description
time
Reports/sets current time (24-hour time)
up
Simulates pressing up pushbutton
zflow
Reports flow through zero air flow controller
avg time
This command reports the photometer averaging time in seconds. The
example below shows that the averaging time is 300 seconds, according to
Table
B–2. This command is only available if the photometer option is
installed.
Send:
avg time
Receive:
avg time 5:300 sec
set avg time selection
This command sets the averaging time according to
example below sets the averaging time to 120 seconds. This command is
only available if the photometer option is installed.
Send:
set avg time 2
Receive:
set avg time 2 ok
Table B–2. Averaging Times
Selection
Averaging Time (seconds)
0
60 seconds
1
90
2
120
3
180
4
240
5
300
gas conc
This command reports the current concentration of gas being supplied by
the Model146i.
Send:
gas conc
Receive:
gas conc actual 100.000 target 1.600 ppm
gflow
zflow
These commands report the current zero air or gas flow.