Setup Procedure
Thermo Fisher Scientific
5. Check for possible damage during shipment.
6. Check that all connectors and circuit boards are firmly attached.
7. Re-install the cover.
Use the following procedure to setup the instrument.
1. Connect a source of zero-air to the inlet port labeled
zero-air source is capable of supplying the full-scale flows of the zero-air
controller at a pressure between 10 and 40 psi.
2. Connect the standard gas cylinders to the ports labeled A, B and C.
3. Install the permeation tube using the following procedure (applicable
only if the permeation option is installed).
a. Remove the instrument cover and locate the permeation oven
("Component Layout"
b. Release the latches on the sides of the oven cover and remove cover.
c. Remove glass chamber assembly
removing) knurled screw, located at the top of the chamber, and
gently pulling assembly upward. Completely remove assembly from
oven.
on
page
7-8).
(Figure
2-3) by loosening (not
Model 146i Instruction Manual
Installation
Setup Procedure
ZERO AIR
. The
2-3