Operation
Service Menu
Set Defaults

Lamp Setting

3-100
Model 146i Instruction Manual
The Set Defaults screen is used to reset the configuration values to factory
defaults.
In the Main Menu, choose Service > Perm Oven Settings > Factory Cal
Oven Therm > Set Defaults.
RESTORE DEFAULT SETTINGS:
FOR PERM OVEN HTR THERM
OPER
MODE
DIAGS ALARM
The Lamp Setting screen is used to adjust the photometer lamp drive. The
intensity and current lamp drive settings are displayed. Adjust the lamp
setting until the intensity is about 10000 Hz. The lamp setting is visible
only if the photometer option is installed and when the calibrator is in
service mode. For more information on the service mode, see
Mode" earlier in the chapter.
In the Main Menu, choose Service > Lamp Setting.
CAL OVEN THERM LOW POINT:
CURRENTLY:
0= 3850 Ohms
SET TO:
MOVE CURSOR
CHANGE VALUE
OPER
MODE
DIAGS ALARM
RESTORE
BENCH LAMP SETTING:
INT:
LAMP SETTING:
CHANGE VALUE
SAVE VALUE
OPER
MODE
DIAGS ALARM
04000
Ohms
SAVE
RESTORE DEFAULT SETTINGS:
FOR PERM OVEN HTR THERM
ARE YOU SURE YOU WANT
PRESS
TO CONFIRM RESTORE
OPER
MODE
DIAGS ALARM
100246 Hz
61.1 %
RESTORE
TO?x
"Service
Thermo Fisher Scientific

Advertisement