Servicing
Analog Output Testing
7-20
Model 146i Instruction Manual
7. Check that the meter is displaying the full scale value. If the meter
reading differs by more than one percent, the analog outputs should be
adjusted. Refer to the "Analog Output Calibration" procedure that
follows.
8. Press
to reset the analog outputs to normal.
Analog
Voltage
Outputs

Figure 7-7. Rear Panel Analog Input and Output Pins

Table 7-4. Analog Output Channels and Rear Panel Pin Connections

Voltage
Pin
Channel
1
14
2
33
3
15
4
34
5
17
6
36
Current Outputs
Analog
Voltage
Inputs
Current
Pin
Channel
1
15
2
17
3
19
4
21
5
23
6
25