Analog Input Calibration - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

Operation
Service Menu
Analog Input
Calibration
Analog Input Calibrate
3-94
Model 146i Instruction Manual
ANALOG OUTPUT CAL:
CONNECT METER TO OUTPUT!
SELECTED OUTPUT:
SET TO:
SAVE VALUE
SET OUTPUT TO:
OPER
The Analog Input Calibration menu is a selection of 8 analog inputs (if the
I/O expansion board option is installed) to calibrate, and allows the user to
select the calibration action zero or span. The analog input calibration is
visible only when the instrument is in service mode. For more information
on the service mode, see
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Analog Input Calibration.
ANALOG INPUT CAL:
>INPUT CHANNEL 1
INPUT CHANNEL 2
INPUT CHANNEL 3
INPUT CHANNEL 4
INPUT CHANNEL 5
INPUT CHANNEL 6
INPUT CHANNEL 7
OPER
ANALOG INPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
OPER
The Analog Input Calibrate Zero screen is used to calibrate the zero state of
the selected analog input.
Zero
INC/DEC
MODE
DIAGS ALARM
"Service
Mode" earlier in the chapter.
MODE
DIAGS ALARM
MODE
DIAGS ALARM
SPAN
V1
3976
10 V