Dynamic gas calibrator

Advertisement

Operation
Service Menu
Calibrate Gas Thermistor
3-96
Model 146i Instruction Manual
The Calibrate Gas Thermistor menu is used for calibrating the permeation
oven gas thermistor using a water bath or known resistance.
In the Main Menu, choose Service > Perm Oven Settings > Cal Gas
Thermistor.
Water Bath
The Calibrate Gas Thermistor Water Bath screen is used to view and set
the permeation oven gas temperature to a known temperature value.
In the Main Menu, choose Service > Perm Oven Settings > Cal Gas
Thermistor > Water Bath.
PERM OVEN SETTINGS:
>CAL GAS THERMISTOR
CAL OVEN THERMISTOR
PERM OVEN SELECTION
FACTORY CAL GAS THERM
FACTORY CAL OVEN THERM
OPER
MODE
DIAGS ALARM
CALIBRATE GAS THERMISTOR:
>WATER BATH
KNOWN RESISTOR
OPER
MODE
DIAGS ALARM
CAL GAS THERM (BATH):
CURRENTLY:
SET TO:
MOVE CURSOR
CHANGE VALUE
OPER
MODE
DIAGS ALARM
o
45.80
C
o
45.00
C
SAVE