Photometer Lamp Temperature - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

Operation
Alarms Menu
Min and Max Photometer
Bench Temperature
Photometer Lamp
Temperature
Min and Max Photometer
Lamp Temperature Limits
3-82
Model 146i Instruction Manual
The Minimum Photometer Bench Temperature alarm limit screen is used
to change the minimum bench temperature alarm limit. The minimum
and maximum bench temperature screens function the same way.
Limits
In the Main Menu, choose Alarms > Phot Bench Temp > Min or Max.
BENCH TEMPERATURE:
ACTUAL MIN:
SET MIN TO:
OPER
The Photometer Lamp Temperature screen displays the current lamp
temperature and sets the minimum and maximum alarm limits. Acceptable
alarm limits range from 45 to 65 °C. If the bench temperature reading goes
beyond either the minimum or maximum limit, an alarm is activated. The
word "ALARM" appears in the Run screen and in the Main Menu.
In the Main Menu, choose Alarms > Phot Lamp Temp.
PHOT LAMP TEMPERATURE:
ACTUAL
>MIN
MAX
OPER
The Minimum Photometer Lamp Temperature alarm limit screen is used
to change the minimum lamp temperature alarm limit. The minimum and
maximum lamp temperature screens function the same way.
In the Main Menu, choose Alarms > Sample Flow > Min or Max.
o
15.0
o
16.0
INC/DEC
SAVE VALUE
MODE
DIAGS ALARM
32.3
50.0
60.0
MODE
DIAGS ALARM
C
C
?
o
C
o
C
o
C