Dynamic gas calibrator

Advertisement

Manual O
GPT/Photometer Control
Thermo Fisher Scientific
Adjust
The Manual screen is used to set the ozonator lamp output manually.
3
In the Main Menu, choose Flow Modes > O3 > Transfer Standard >
Manual.
The ozonator output level can be set using the Ozonator GPT Select
screen. Alternately, if Photometer Control is selected then all ozonator
levels are shown in parts per million (ppm) instead of percent (%). The
MANUAL screen is used to set the ozonator output to the manual level, in
ppm or %.
In the Main Menu, choose Flow Modes > O3 > GPT/Photometer
Control > Level 1-5.
O3 LEVEL SELECTION:
>LEVEL1
LEVEL2
LEVEL3
LEVEL4
LEVEL5
MANUAL
OPER
MODE
DIAGS ALARM
FLOW MODE O3 LEVEL 1:
O3 LEVEL:
TFLOW:
SAVING
SAVE FLOW MODE
OPER
MODE
DIAGS ALARM
MANUAL O3 LAMP SETTING:
CURRENTLY:
SET TO:
CHANGE VALUE
SAVE VALUE
OPER
MODE
DIAGS ALARM
100.0 %
80.0 %
50.0 %
20.0 %
10.0 %
0.000 %
20.0 %
5000 SCCM
0.0 %
5.0 % ?
O3 LEVEL SELECTION:
>LEVEL1
LEVEL2
LEVEL3
LEVEL4
LEVEL5
MANUAL
OPER
MODE
DIAGS ALARM
Model 146i Instruction Manual
Operation
Flow Modes
1.000 PPM
0.800 PPM
0.500 PPM
0.200 PPM
0.100 PPM
0.000 PPM
3-17