Photometer Bench Temperature - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

Min and Max Pressure
Photometer Bench
Temperature
Thermo Fisher Scientific
PRESSURE:
ACTUAL
>MIN
MAX
OPER
The Minimum Pressure alarm limit screen is used to change the minimum
Limits
temperature alarm limit. The minimum and maximum pressure screens
function the same way.
In the Main Menu, choose Alarms > Pressure > Min or Max.
PRESSURE:
ACTUAL MIN:
SET MIN TO:
OPER
The Photometer Bench Temperature screen displays the current bench
temperature and sets the minimum and maximum alarm limits. Acceptable
alarm limits range from 10 to 50 °C. If the bench temperature reading goes
beyond either the minimum or maximum limit, an alarm is activated. The
word "ALARM" appears in the Run screen and in the Main Menu.
In the Main Menu, choose Alarms > Phot Bench Temp.
PHOT BENCH TEMPERATURE:
ACTUAL
>MIN
MAX
OPER
746.9 mmHg
200.0 mmHg
1000.0 mmHg
MODE
DIAGS ALARM
200.0 mmHg
250.0 mmHg ?
INC/DEC
SAVE VALUE
MODE
DIAGS ALARM
32.3
15.0
45.0
MODE
DIAGS ALARM
Alarms Menu
o
C
o
C
o
C
Model 146i Instruction Manual
Operation
3-81