Diagnostics; Table B-3. Alarm Trigger Values - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

C-Link Protocol Commands

Diagnostics

Diagnostics
B-14
Model 146i Instruction Manual
Table B–3. Alarm Trigger Values
Value
Alarm Trigger
0
Floor
1
Ceiling
diag volt mb
This command reports the diagnostic voltage measurements on the
motherboard. The sequence of voltages is: Positive 24, positive 15, positive
5, positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag volt mb
Receive:
diag volt mb 24.1 14.9 4.9 3.2 -3.2
diag volt mib
This command reports the diagnostic voltage measurements on the
measurement interface board. The sequence of voltages is: Positive 24,
positive 15, negative 15, positive 5, positive 3.3. Each voltage value is
separated by a space.
Send:
diag volt mib
Receive:
diag volt mib 24.1 14.9 -14.9 4.9 3.2
diag volt iob
This command reports the diagnostic voltage measurements on the I/O
expansion board. The sequence of voltages is: Positive 24, positive 5,
positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag volt iob
Receive:
diag volt iob 24.1 4.9 3.2 -3.2
event event params
event = 1-32
This command reports the parameters programmed for the specified event
event.
Send:
event 1 params
Receive:
event 1 params gas A span 1 ozon 3 perm 1 photo
internal
event status
This command reports the current event that is being executed 1-32. A 0
response indicates no event is currently running.
Send:
event status
Receive:
event status 1