Operation
Instrument Controls Menu
3-68
Model 146i Instruction Manual
Table Point
The Table Point submenu allows the user to set up an individual table
point.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Channel > Point 1-10.
Volts
The Volts screen allows the user to set the input voltage for the selected
table point in the conversion table, from 0.00 to 10.50. The default table is
a two-point table with point 1: 0.00 V = 000.0 U and point 2: 10.00 V =
10.0 U.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Channel > Select Point > Volts.
User Value
The User Value screen allows the user to set the output value for the
corresponding input voltage for the selected table point in the conversion
table, from -9999999 to 99999999. The default table is a two-point table
with point 1: 0.00 V = 000.0 U and point 2: 10.00 V = 10.0 U.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Table Point > User Value.
TABLE POINT 01 CONFIG:
>VOLTS
USER VALUE
OPER
MODE
DIAGS ALARM
TABLE POINT 01 VOLTS:
CURRENTLY:
SET TO:
MOVE CURSOR
CHANGE VALUE
OPER
MODE
DIAGS ALARM
0.00
0.00
0.00
00.00
SAVE