Dynamic gas calibrator

Advertisement

Operation
Instrument Controls Menu
Streaming Data
Configuration
Streaming Data Interval
3-50
Model 146i Instruction Manual
COMMUNICATION PROTOCOL:
CURRENTLY:
SET TO:
OPER
MODE
The Streaming Data Configuration menu is used to allow for configuration
of the 8 streaming data output items, streaming interval, current data
format, and current timestamp setting. The Choose Item Signal submenu
displays a list of the analog output signal group choices to choose from.
Choices are Concentrations, Other Measurements, and Analog Inputs (if
the I/O expansion board option is installed).
In the Main Menu, choose Instrument Controls > Communication
Settings > Streaming Data Config.
STREAMING DATA CONFIG
>INTERVAL
ADD LABELS
PREPEND TIMESTAMP
ITEM 1
ITEM 2
ITEM 3
ITEM 4
OPER
MODE
CHOOSE STREAM DATA:
>CONCENTRATIONS
OTHER MEASUREMENTS
ANALOG INPUTS
OPER
MODE
The Streaming Data Interval screen is used to adjust the streaming data
interval. The following interval times are available: 1, 2, 5, 10, 20, 30, 60,
90, 120, 180, 240, and 300 seconds.
In the Main Menu, choose Instrument Controls > Communication
Settings > Streaming Data Config > Streaming Data Interval.
CLINK
STREAMING ?
CHANGE VALUE
SAVE VALUE
DIAGS ALARM
10 SEC
NO
YES
CONC
TGFLO
AGFLO
TZFLO
DIAGS ALARM
DIAGS ALARM