Manual Flow

Thermo Fisher Scientific
GAS A SPAN 1 FLOW:
OPER
The Manual Flow screen displays the zero-air and gas flows, and lets you
manually set the zero-air and gas flows.
In the Main Menu, choose Gas Setup > Gas A-F > Manual.
GAS A MANUAL FLOWS:
>ZERO AIR
GAS
OPER
Zero Air
The Manual Zero Air Flow screen sets the manual zero-air flow value for
the previously selected gas. The screen identifies the minimum and
maximum acceptable zero-air flow rates and lets the user select a flow rate
between these two settings. If the selected zero-air flow is lower or higher
than the acceptable range, "TOO LOW" or TOO HIGH" appears on the
screen.
In the Main Menu, choose Gas Setup > Gas A-F > Manual > Zero Air.
MAN GAS A ZERO AIR FLOW:
OPER
MINIMUM:
444 SCCM
MAXIMUM:
01000 SCCM
SET TO:
00800
MOVE CURSOR
CHANGE VALUE
MODE
DIAGS ALARM
08000 CCM
030.00 CCM
MODE
DIAGS ALARM
MINIMUM:
00400 SCCM
MAXIMUM:
10000 SCCM
SET TO:
01001
MOVE CURSOR
CHANGE VALUE
MODE
DIAGS ALARM
SCCM ?
SAVE
SCCM ?
SAVE
Model 146i Instruction Manual