C-Link Protocol Commands
Datalogging
B-20
Model 146i Instruction Manual
Send:
lrec layout
Receive:
lrec layout %s %s %lx %f %f %f %f
t D L ffff
flags conc tgflo agflo tzflo azflo ttflo atflo pres
lrec mem size
srec mem size
These commands report the long and short records that can be stored with
the current settings and the number of blocks reserved for long and short
records. To calculate the number of short records per block, add 2 to the
number of records, and then divide by the number of blocks. The example
below shows that 10 blocks were reserved for long records, and the
maximum number of long records that can be stored in memory is 2038.
Send:
lrec mem size
Receive:
lrec mem size 2038 recs, 10 blocks
lrec per
srec per
These commands report the long and short records logging period. The
example below shows that the short record logging period is 5 minutes.
Send:
srec per
Receive:
srec per 5 min
set srec per value
set srec per value
value = | 1 | 5 | 15 | 30 | 60 |
These commands set the long and short records logging period to value in
minutes. The example below sets the long record logging period to 15
minutes.
Send:
set lrec per 15
Receive:
set lrec per 15 ok
no of lrec
no of srec
These commands report the number of long and short records stored in the
long and short records memory. The example below shows that 50 long
records have been stored in the memory.
Send:
no of lrec
Receive:
no of lrec 50 recs
Thermo Fisher Scientific